High resolution corneal and single pulse imaging with line field spectral domain optical coherence tomography

Articolo
Data di Pubblicazione:
2016
Abstract:
We report the development of a Spectral Domain Line Field Optical Coherence Tomography (LF-OCT) system, using a broad bandwidth and spatial coherent Super-Continuum (SC) source. With conventional quasi-Continuous Wave (CW) setup we achieve axial resolutions up to 2.1 μm in air and 3D volume imaging speeds up to 213 kA-Scan/s. Furthermore, we report the use of a single SC pulse, of 2 ns duration, to temporally gate an OCT B-Scan image of 70 A-Scans. This is the equivalent of 35 GA-Scans/s. We apply the CW setup for high resolution imaging of the fine structures of a human cornea sample ex-vivo. The single pulse setup is applied to imaging of a coated pharmaceutical tablet. The fixed pattern noise due to spectral noise is removed by subtracting the median magnitude A-Scan. We also demonstrate that the Fourier phase can be used to remove aberration caused artefacts.
Tipologia CRIS:
1.1 Articolo in rivista
Elenco autori:
Lawman, S.; Dong, Y.; Williams, B. M.; Romano, V.; Kaye, S.; Harding, S. P.; Willoughby, C.; Shen, Y. -C.; Zheng, Y.
Autori di Ateneo:
ROMANO VITO
Link alla scheda completa:
https://iris.unibs.it/handle/11379/571489
Pubblicato in:
OPTICS EXPRESS