Data di Pubblicazione:
2019
Abstract:
We report the first-in-human retrieval of a chronically implanted Micra transcatheter pacing system (TPS; Medtronic, Minneapolis, MN, USA). A 41-year-old woman suffering from third-degree atrioventricular block was successfully implanted with a TPS in the low septum of the right ventricle. After 29 months, due to an increased electrical threshold and 100% pacing percentage, the device showed low battery voltage. Despite the long implant duration, the attempt of device retrieval using a snare loop inserted in the delivery system of a TPS was successful. In the same procedure, a new leadless pacemaker was implanted in the high right ventricular septum with optimal pacing threshold.
Tipologia CRIS:
1.1 Articolo in rivista
Keywords:
end-of-life management; leadless pacemaker; leadless pacemaker retrieval; micra transcatheter pacing system;
Elenco autori:
Curnis, A.; Cerini, M.; Mariggio, D.; Mitacchione, G.; Giacopelli, D.; Inama, L.; Bontempi, L